研究目的
To apply a high spectral resolution optical feedback cavity enhanced absorption spectroscopy (OF-CEAS) for ammonia detection in exhaled human breath, aiming for disease diagnosis and health status monitoring.
研究成果
The OF-CEAS system demonstrated high sensitivity and accurate discrimination for NH3 detection in exhaled human breath, with potential applications in clinical diagnostics and atmospheric sciences. The system's compactness and robustness make it suitable for in-situ measurements.
研究不足
The study is limited by the need for low sample pressure to eliminate interference from other components in exhaled human breath, and the spectral resolution is dependent on the laser frequency tuning and cavity dither.
1:Experimental Design and Method Selection
The study utilizes OF-CEAS with a fibered DFB diode laser locked to a V-shaped cavity for high sensitivity and spectral resolution measurement of NH3 in exhaled human breath.
2:Sample Selection and Data Sources
Exhaled breath samples from healthy subjects and patients suffering from Helicobacter pylori infection were collected into Tedlar bags and analyzed.
3:List of Experimental Equipment and Materials
Fibered DFB laser diode, V-shaped cavity, photodetectors, optical switch, PM fiber couplers, pressure regulators, and a dry pump.
4:Experimental Procedures and Operational Workflow
Samples were exhaled into Tedlar bags, filtered, and flowed through the V-shaped cavity at a controlled rate and pressure. The absorption spectra were recorded and analyzed.
5:Data Analysis Methods
The absorption coefficient spectra were converted from cavity transmission signals using a specific formula, and Voigt profile fitting was applied to the experimental absorption lines.
独家科研数据包,助您复现前沿成果,加速创新突破
获取完整内容-
Universal laser diode mount
LM14S2
Thorlabs
Houses the DFB laser diode.
-
Current deriver
LD202C
Thorlabs
Regulates the injection current of the laser diode.
-
Temperature controller
TED200C
Thorlabs
Regulates the temperature of the laser diode.
-
Photodetector
PDA400
Thorlabs
Detects optical intensity signals.
-
High precision wavemeter
WA-1500
EXFO
Calibrates the wavenumber corresponding to the injection current.
-
DFB laser diode
SWLD-1531.5-15-PM
Allwave Lasers
Light source for the OF-CEAS system, emitting at 1531.6 nm.
-
Dynamic Signal Acquisition and Generation device
NI-PCI-4461
National Instruments
Generates voltage ramp for laser frequency scanning and records signals.
-
Tedlar bag
Collects exhaled breath samples.
-
Midisart-2000 air filter
Sartorius
Removes air particulates and offers online sterilization.
-
Ultra-high purity in-line gas filter
Prevents organic macromolecules from entering the cavity.
-
登录查看剩余8件设备及参数对照表