To evaluate the effectiveness of the Nd:YAP laser in the treatment of dentin hypersensitivity, compare the temperature rise during laser irradiation at three different dentine thicknesses, and analyse the composition of the dentine-lased surface.
研究成果
The Nd:YAP laser at a power density of 356 W/cm2 (1.4 W average power) is effective in treating dentine hypersensitivity by occluding dentinal tubules without significant changes in the element distribution and is safe for the pulp when used within the tested parameters.
研究不足
The study is an ex vivo model, which may not fully replicate in vivo conditions. The sample size is relatively small, and the study does not assess long-term effects of the treatment.
1:Experimental Design and Method Selection:
The study used an ex vivo design with human molars divided into groups for different treatments including EDTA irrigation and Nd:YAP laser irradiation at varying power densities.
2:Sample Selection and Data Sources:
33 extracted caries- and restoration-free human molars were selected.
3:List of Experimental Equipment and Materials:
Nd:YAP laser (LOKKI Dt, LOBEL Medical, France), scanning electron microscope (JSM-5310LV35, Jeol Ltd, Tokyo, Japan), thermocouples (Pico Technology, TC-08 USB, United Kingdom).
4:Experimental Procedures and Operational Workflow:
Teeth were sectioned, treated with EDTA or laser at different parameters, and observed under SEM. Temperature changes were measured at different dentine thicknesses.
5:Data Analysis Methods:
Statistical analysis was performed using GraphPad Prism Statistic Software using an analysis of variance (ANOVA) test for multiple comparisons.