研究目的
Investigating the application of the AC Quantum Voltmeter for impedance ratio measurements at voltages up to 5 V and frequencies up to 1 kHz.
研究成果
The results show that for Fluke 5720A, standard deviation is limited due to the phase-lock principle applied. The output voltage of Aivon DualDAC2 is more stable with lower noise, but only with no change in connection, load or frequency.
研究不足
The output voltage of Aivon DualDAC2 is more stable with lower noise but only with no change in connection, load or frequency.
1:Experimental Design and Method Selection:
The AC Quantum Voltmeter measures the RMS voltages on each resistor in a time sequence.
2:Sample Selection and Data Sources:
Two self-developed resistors of nominal values of 10 kΩ are used.
3:List of Experimental Equipment and Materials:
A calibrator Fluke 5720A and an Aivon DualDAC2 source are used as the AC voltage source.
4:Experimental Procedures and Operational Workflow:
The impedances are measured in a potentiometric way.
5:Data Analysis Methods:
The ratio R1/R2 is calculated as deviation in ppm from nominal value of 1.