研究目的
To report the first-in-human images with a hybrid imaging catheter combining intravascular ultrasound (IVUS) and optical coherence tomography (OCT) for characterizing coronary atherosclerosis and coronary interventions.
研究成果
The first in vivo experience of hybrid IVUS and OCT demonstrates the synergistic ability of these modalities to characterize coronary atherosclerosis and coronary interventions, with OCT providing better near-field findings and IVUS offering deeper feature assessment.
研究不足
The study is limited by being a first-in-human experience with a small sample size (n=1). Further studies are needed to validate the findings and assess the catheter's performance in a larger population.
1:Experimental Design and Method Selection:
The study utilized a novel hybrid IVUS and OCT catheter for imaging coronary atherosclerosis and stents in a human patient. The catheter features a 2.8-F imaging window with a 40-MHz center frequency for IVUS imaging, with IVUS and OCT beams aligned for coregistration and simultaneous image review.
2:8-F imaging window with a 40-MHz center frequency for IVUS imaging, with IVUS and OCT beams aligned for coregistration and simultaneous image review.
Sample Selection and Data Sources:
2. Sample Selection and Data Sources: Imaging was performed in a 74-year-old male patient with a recent inferior ST-segment elevation myocardial infarction treated with primary percutaneous coronary intervention.
3:List of Experimental Equipment and Materials:
The hybrid IVUS and OCT catheter, Promus 2.75 × 20 mm stent (Boston Scientific, Natick, Massachusetts), and contrast flush were used.
4:75 × 20 mm stent (Boston Scientific, Natick, Massachusetts), and contrast flush were used.
Experimental Procedures and Operational Workflow:
4. Experimental Procedures and Operational Workflow: Imaging was performed over 90 mm of the right coronary artery at a pull-back speed of 25 mm/s, with contrast flushing at 3 ml/s for 5 s. Post-stent implantation imaging was performed over a 70-mm length at a pull-back speed of 25 mm/s and contrast flush at 4 ml/s for 4 s.
5:Data Analysis Methods:
The complementary roles of IVUS and OCT were analyzed to characterize atherosclerotic plaque types and assess stent apposition and dissection.