研究目的
To develop a high-precision Compton imaging technique for 4.4 MeV prompt gamma-rays to monitor proton therapy online, addressing the limitations of current PET methods in accurately matching the proton dose distribution.
研究成果
The developed Compton camera demonstrated high precision in imaging 4.4 MeV prompt gamma-rays, with the image sharply concentrated on the Bragg peak as expected. This technique shows promise for online monitoring of proton therapy, offering a more accurate match to the proton dose distribution than PET.
研究不足
The study was conducted with a 70 MeV proton beam, and further experiments are needed with higher energy beams (200 MeV) and more complex phantoms to fully assess clinical applicability.
1:Experimental Design and Method Selection:
Developed a novel, 3-D position sensitive Compton camera based on Ce:GAGG scintillators coupled with MPPC arrays, optimized for imaging in the 1-10 MeV range.
2:Sample Selection and Data Sources:
Used a 70 MeV proton beam irradiated on a PMMA phantom to mimic clinical beam environments.
3:List of Experimental Equipment and Materials:
Ce:GAGG scintillators, MPPC arrays, tungsten shields, BGO shields.
4:Experimental Procedures and Operational Workflow:
Positioned the Compton camera at a distance from the phantom, applied energy cut conditions for event selection, and used the list-mode maximum likelihood expectation maximization method for image reconstruction.
5:Data Analysis Methods:
Applied methods to discriminate multiple-Compton and escape events to reduce imaging noise.