研究目的
To develop a novel array detector for overcoming the dosimetry challenges of measuring in very short pulsed charged particle beams, specifically addressing the issues of recombination and polarity corrections in ion chamber measurements.
研究成果
The initial measurements confirmed the validity of the FLUKA geometry and beam parameters used in the simulations, with a 5% difference between experimental and simulated calibration factors within the error margin. This validation allows for further investigation into the reciprocal influences of the chambers within the array detector.
研究不足
The study is limited by the initial inhomogeneity of the proton beam and the need for corrections due to the low energy of the protons. The reciprocal influences of the ion chambers in the array detector also present a challenge that requires further investigation.
1:Experimental Design and Method Selection:
The study proposes an array detector with four identical ion chambers for simultaneous measurement of recombination and polarity corrections. FLUKA simulations are used to model the detector's response.
2:Sample Selection and Data Sources:
Measurements were performed with an Advanced MarkusTM chamber in a 3-MeV proton beam from a TandetronTM accelerator.
3:List of Experimental Equipment and Materials:
PTW Advanced Markus type 34 045 ion chamber, PTW Unidos E electrometer, Gafchromic? EBT2 and HD-V2 films, EPSON EXPRESSION 11 000XL scanner, PTW Mephysto Mcc software.
4:Experimental Procedures and Operational Workflow:
Initial measurements were conducted to validate FLUKA simulations, including beam homogeneity improvement using a gold foil.
5:Data Analysis Methods:
Data analysis involved comparing experimental calibration factors with FLUKA simulation results, using statistical techniques and software tools for analysis.