研究目的
Investigating the process monitoring of virus-like particle reassembly by diafiltration using UV/Vis spectroscopy and light scattering to improve understanding and control of VLP assembly in biopharmaceutical production.
研究成果
The developed CFF setup with on-line UV/Vis and light scattering sensors effectively monitors VLP reassembly, providing real-time data on assembly progress, protein concentration, and structural changes. Evidence of aggregate-related inhibition was found, and PLS models enabled accurate VLP concentration prediction. The setup enhances process understanding and control, with potential for improved product quality and process intensification in biopharmaceutical production.
研究不足
The study is limited to specific HBcAg variants and may not generalize to other VLPs. The on-line DLS measurements are affected by convective flow, potentially underestimating particle sizes. Membrane clogging occurred in some runs, limiting data completeness. Off-line SEC may not detect large aggregates, and the PLS models require calibration for each variant.
1:Experimental Design and Method Selection:
A custom-made Cross-Flow Filtration (CFF) setup with an on-line measurement loop was developed for Process Analytical Technology (PAT) implementation. The setup included UV/Vis spectroscopy and light scattering (DLS and SLS) for real-time monitoring of protein concentration, tertiary structure, and quaternary structure during VLP reassembly.
2:Sample Selection and Data Sources:
Three chimeric Hepatitis B core Antigen (HBcAg) variants (VLP A, B, and C) provided by BioNTech Protein Therapeutics GmbH were used. Stock solutions were in disassembly buffer (
3:5 M urea, 50 mM Tris, pH 0) and diluted to 1 g/L protein concentration before experiments. List of Experimental Equipment and Materials:
Equipment included a KrosFlo KRIIi CFF unit with mPES hollow fiber membrane module (10 kDa cutoff, 13 cm2 area), Gilson Minipuls 3 peristaltic pump, Zetasizer Nano ZSP photometer with flowcell, Ultimate DAD-3000 Diode Array Detector with flowcell, FR-902 flow restrictor, NI USB-6008 data acquisition device, and various filters and buffers. Materials included HBcAg variants, urea, Tris, potassium phosphate, and ultrapure water.
4:Experimental Procedures and Operational Workflow:
The CFF unit and measurement loop were filled with water, equilibrated, then flushed with disassembly buffer and loaded with protein solution. Diafiltration was initiated at TMPs of
5:25, 5, or 1 bar with reassembly buffer. On-line sensors monitored continuously, and samples were taken every 5 diafiltration volumes for off-line analysis (SEC, DLS, TEM). Data Analysis Methods:
Data was acquired and synchronized using a custom Matlab application. UV/Vis spectra were processed with Savitzky-Golay filters for second derivatives to assess tyrosine and tryptophan solvatization. DLS provided z-average diameters, SLS provided scattered-light intensity. PLS models were calibrated for VLP concentration prediction using cross-validation.
独家科研数据包,助您复现前沿成果,加速创新突破
获取完整内容-
Zetasizer Nano ZSP photometer
Nano ZSP
Malvern Instruments
Used for dynamic light scattering (DLS) and static light scattering (SLS) measurements to monitor particle size and intensity.
-
Ultimate 3000 RS UHPLC system
3000 RS
Thermo Fisher Scientific
High-performance liquid chromatography system for SEC analysis, including pump, autosampler, column compartment, and detector.
-
Titan3 80-300 microscope
Titan3 80-300
FEI Company
Transmission electron microscope for imaging VLPs to verify assembly.
-
KrosFlo KRIIi CFF unit
KRIIi
Spectrum Labs
Core unit for cross-flow filtration experiments, used for VLP reassembly by diafiltration.
-
mPES hollow fiber membrane module
Spectrum Labs
Membrane for filtration in the CFF unit, allowing buffer exchange and concentration.
-
Gilson Minipuls 3 peristaltic pump
Minipuls 3
Gilson
Pump in the on-line measurement loop for fluid circulation.
-
Ultimate DAD-3000 Diode Array Detector
DAD-3000
Dionex Corporation
UV/Vis spectrometer for acquiring absorption spectra to monitor protein concentration and structural changes.
-
FR-902 flow restrictor
FR-902
GE Healthcare
Sets minimal back pressure in the measurement loop to reduce fluctuations and air bubbles.
-
NI USB-6008 data acquisition device
USB-6008
National Instruments
Controls the pump of the on-line measurement loop and acquires data from sensors.
-
Minisart glass fiber syringe filter
Minisart
Sartorius Stedim Biotech
Filter in the on-line loop to retain bubbles and large particles, improving measurement quality.
-
Topolino magnetic stirrer
Topolino
IKA Werke
Ensures homogeneous mixing of the retentate reservoir in the CFF setup.
-
Centrifuge 5810R
5810R
Eppendorf
Used for centrifuging samples to settle large particles before SEC analysis.
-
Sepax SRT SEC-1000 column
SRT SEC-1000
Sepax Technologies
Size-exclusion chromatography column for off-line analysis of VLP and aggregate concentrations.
-
登录查看剩余11件设备及参数对照表