修车大队一品楼qm论坛51一品茶楼论坛,栖凤楼品茶全国楼凤app软件 ,栖凤阁全国论坛入口,广州百花丛bhc论坛杭州百花坊妃子阁

oe1(光电查) - 科学论文

1 条数据
?? 中文(中国)
  • Re-evaluation of the correction factors for the GROVEX

    摘要: The GROVEX (GROssVolumige EXtrapolationskammer, large-volume extrapolation chamber) is the primary standard for the dosimetry of low-dose-rate interstitial brachytherapy at the Physikalisch-Technische Bundesanstalt (PTB). In the course of setup modifications and re-measuring of several dimensions, the correction factors have been re-evaluated in this work. The correction factors for scatter and attenuation have been recalculated using the Monte Carlo software package EGSnrc, and a new expression has been found for the divergence correction. The obtained results decrease the measured reference air kerma rate by approximately 0.9% for the representative example of a seed of type Bebig I25.S16C. This lies within the expanded uncertainty (k = 2).

    关键词: dosimetry,ionisation chamber,extrapolation chamber,low-dose rate brachytherapy,primary standard

    更新于2025-09-09 09:28:46