研究目的
To evaluate the influence of out-of-plane motions on the performance of axial strain and axial shear strain estimations in two-dimensional noninvasive vascular ultrasound elastography.
研究成果
The performance of the LSME for axial strain and shear estimations is robust in the presence of out-of-plane motion magnitudes corresponding to translations of carotid artery plaques observed clinically, even with motions up to 2 mm.
研究不足
The study is limited by the complexity of human tissue structure of carotid plaques observed clinically, which may affect the correlation coefficients differently than in vitro experiments. Additionally, the study focuses on out-of-plane motions up to 3 mm, which may not cover all possible clinical scenarios.
1:Experimental Design and Method Selection:
Designed a diseased carotid bifurcation phantom with a 70% stenosis and an in vitro experimental setup to simulate orthogonal out-of-plane motions. The Lagrangian speckle model estimator (LSME) was used to estimate axial strains and shears under pulsatile conditions.
2:Sample Selection and Data Sources:
Used a clinical LSME dataset obtained from 18 participants with carotid artery stenosis to verify findings of in vitro experiments.
3:List of Experimental Equipment and Materials:
Diseased carotid bifurcation phantom, Verasonics research scanner (V1-128), L7-4 probe, Sonix ES500 RP ultrasonic system, L14-5/38 linear array probe.
4:Experimental Procedures and Operational Workflow:
Simulated out-of-plane motions of 1 mm, 2 mm, and 3 mm. Acquired ultrafast plane wave data and performed NIVE on clinical dataset.
5:Data Analysis Methods:
Used correlation coefficients and normalized cross-correlations to evaluate the influence of out-of-plane motions on strain and shear estimations.
独家科研数据包,助您复现前沿成果,加速创新突破
获取完整内容-
Pulsatile pump
1421
Harvard Apparatus
Used to vary intra-luminal pressure in the phantom.
-
Verasonics research scanner
V1-128
Verasonics Inc.
Used to transmit and receive plane waves for ultrafast plane wave data acquisition.
-
Linear array transducer
ATL L7-4
Philips
Used for ultrasound imaging with 128 elements.
-
Sonix ES500 RP ultrasonic system
Sonix ES500 RP
Ultrasonix Medical Corp.
Used for clinical ultrasound imaging.
-
Linear array probe
L14-5/38
Ultrasonix Medical Corp.
Used for clinical ultrasound imaging with 128 elements.
-
Mini-shaker
4810
Brüel and Kj?r
Used to generate out-of-plane motions.
-
Power amplifier
2706
Brüel and Kj?r
Used to adjust magnitudes of vibration for the mini-shaker.
-
Sonometricmetry system
Sonometrics Corporation
Used to calibrate magnitudes of out-of-plane motions.
-
登录查看剩余6件设备及参数对照表