研究目的
To measure the extremely small base angle of an axicon using phase-shifting digital holography.
研究成果
The phase-shifting digital holography method can accurately measure the base angle of an axicon, with errors of 1.94%, 4.43%, and 1.63% for axicons with base angles of 0.5°, 1°, and 1°, respectively.
研究不足
The errors in measurement may be caused by the axicon not being well-manufactured and the phase-shift introduced by SLM deviating from the required value.
1:Experimental Design and Method Selection:
The phase-shifting digital holography technique was applied to measure the base angle of axicons. The method involves recording interference holograms with different phase shifts introduced by a spatial light modulator (SLM).
2:Sample Selection and Data Sources:
Three axicons with base angles of
3:5°, 1°, and 1° were measured. List of Experimental Equipment and Materials:
A laser beam, beam expander, SLM, polarizer, axicon, beam splitter (BS3), and charge coupled device (CCD).
4:Experimental Procedures and Operational Workflow:
A laser beam is split into two beams; one is reflected onto the SLM, and the other passes through the axicon. The two beams are combined and interfered, with results recorded by a CCD.
5:Data Analysis Methods:
The phase distribution was solved using four, five, or eight phase-shifted holograms, with the best results obtained using eight holograms.