研究目的
Investigating the potential of supercontinuum laser sources to enhance Fourier transform infrared (FT-IR) spectroscopy by replacing conventional thermal emitters, focusing on noise levels, stability, and analytical performance.
研究成果
The study demonstrated that FT-IR spectrometers equipped with novel broadband mid-IR supercontinuum lasers could outperform traditional systems, providing superior performance such as extended interaction path lengths while maintaining low noise levels. A four-times-enhanced detection limit was achieved for aqueous formaldehyde solutions.
研究不足
The study was limited by the dynamic range of the detector when measuring over the entire mid-IR part of the emission spectrum, requiring the use of filters or smaller aperture diameters.
1:Experimental Design and Method Selection:
The study introduced a supercontinuum laser source as a replacement for conventional thermal emitters in FT-IR spectroscopy, proposing an approach to efficient coupling of pulsed mid-IR supercontinuum sources to FT-IR spectrometers.
2:Sample Selection and Data Sources:
Aqueous formaldehyde solutions were used to test the analytical performance of the supercontinuum-based FT-IR spectrometer.
3:List of Experimental Equipment and Materials:
A commercial FT-IR instrument (Bruker Optics, Vertex 70) and a pulsed supercontinuum laser (NKT Photonics, SuperK) were used.
4:Experimental Procedures and Operational Workflow:
The study examined pulse-to-pulse energy fluctuations, system performance, noise, and long-term stability. Comparative measurements were performed with a conventional FT-IR instrument.
5:Data Analysis Methods:
The analytical performance was evaluated by comparing detection limits for aqueous formaldehyde solutions using both the supercontinuum-based and conventional systems.
独家科研数据包,助您复现前沿成果,加速创新突破
获取完整内容-
Supercontinuum laser source
SuperK
NKT Photonics
Replacement for conventional thermal emitters in FT-IR spectroscopy
暂无现货
预约到货通知
-
FT-IR spectrometer
Vertex 70
Bruker Optics
Spectroscopic analysis using supercontinuum laser source
暂无现货
预约到货通知
-
Bandpass filter
FB3500-500
Thorlabs
Selection of the investigated spectral range
-
Power meter
LM-10 HTD
Coherent
Measurement of the average power of the transmitted band
暂无现货
预约到货通知
-
Lock-in amplifier
SR844
Stanford Research Systems
Integration of the pulsed signal to suppress its pulsed nature
暂无现货
预约到货通知
-
Monochromator
GM500
Gilden Photonics
Characterization and verification of pulse-to-pulse energy fluctuations
暂无现货
预约到货通知
-
High-speed MCT detector
PVM series
Vigo
Detection of supercontinuum laser pulses
暂无现货
预约到货通知
-
Liquid flow cell
PIKE Technologies
Sample insertion for spectroscopic analysis
暂无现货
预约到货通知
-
登录查看剩余6件设备及参数对照表
查看全部